Test Method
Carefully read the reagent instruction before using the test kit and strictly operate
according to the instruction to ensure reliable results. Bring all reagents to room
temperature (18-25℃) before use.
(1) Startup: Click “STD Mode” in the main menu to enter the measurement interface,
click “Item” to select the desired test item and click “Type” to select the sample type.
(2) Click “Lot No.” to enter the card swiping interface, place magcard of the
corresponding item to the magnetic induction zone, when hearing a “di” sound, the
magcard is swiped successfully, check whether the magcard and the test card are of
the same batch. (Note: reagents are precalibrated and specific calibration curve
parameters for each batch of reagents have been stored in the magcard.)
(3) Remove the test sample and required reagents from storage conditions and
equilibrate to room temperature.
(4) Remove the test card from the packaging bag and lay it flat on a dry surface.
(5) Sampling:
(a) Samples: Take 40 μL of samples and add them to the sample diluent (160μL),
mix thoroughly for 30s. Take 100 μL diluted sample, drop vertically to the sample port
on the test card directly and start timing.
(b) After adding the sample, the positive sample can be detected within 15 minutes, it
is recommended that the final test results be recorded and recorded within 15-20
minutes.